Technical guides

Technical guide

Laser Beam Profiler Selection and Measurement Guide

By
Shinhotek LUMOS
Published
2026-11-19

What to check when you introduce a laser beam profiler or replace an existing one, in order: principle, beam types, model selection, measurement environment, software and procedure.

1. What a beam profiler does

A beam profiler measures the intensity distribution across a laser beam's cross-section and reports:

  • Beam diameter (x and y) and ellipticity
  • Centroid position and its drift over time
  • Distribution shape such as uniformity and flatness

A power meter tells you how strong the beam is; a beam profiler tells you where the energy is and how it is spread. Laser process quality mostly depends on spot size and shape at the work plane, so the two are used together.

2. Principle: camera-based profiling and ISO 11146

LUMOS profilers are camera-based. The beam is attenuated and imaged directly onto a CMOS sensor, so the full 2D profile is captured in one frame. Pulsed lasers are measured by synchronizing exposure with the trigger input.

Beam diameter follows the ISO 11146 second-moment (D4σ) definition, which accounts for the whole distribution and lets beams of different shapes be compared on the same basis. Lumosity also calculates Knife-Edge, Slit, Ellipse Fit, Threshold and Min Area diameters.

Three conditions for reliable results:

  • The beam covers at least 20 pixels on the sensor
  • The sensor is not saturated: set ND attenuation and exposure so the peak stays within range
  • Background is subtracted: subtract a dark frame taken with the beam off, or the diameter is inflated

3. Different beam types, different figures

Beam typeKey figuresWatch out for
GaussianD4σ (= 1/e²) diameter, ellipticity, centroidSaturation makes the diameter look larger
Top-hat (flat-top)Plateau size, uniformity, edge steepnessA single D4σ does not describe the shape
BesselCore diameter, energy share in the ringsWhole-profile D4σ is much larger than the core
Multi-mode, multi-spotPosition and intensity per spotSplit the image into regions of interest

The separate guide "Gaussian and Top-hat Beam Measurement Comparison" covers the first two in detail.

4. Choosing a model: wavelength, beam size, measurement plane

Decide the wavelength, the beam size to measure and where you measure (raw beam or focused spot); that largely determines the model.

ModelUseWavelengthMinimum spot
LUM-B-D (LUM-B-DUV)Raw beam, incl. DUV/UV200–1064 nmØ 60 µm
LUM-B-NRaw beam343–1064 nmØ 60 µm
LUM-B-MRaw beam, larger sensor (14.4 × 9.9 mm)343–1064 nmØ 100 µm
LUM-B-L / LUM-B-L-UVLarge beams (38.75 × 27.75 mm sensor, 51 MP)400–1064 / 343–1064 nmØ 92 µm
LUM-FFocused spots (×20, ×10, ×05 objectives)240–1064 nm by modelØ 3 µm (×20)
LUM-Z3D distribution around focus, spot size and depth of focus343–1064 nm> 3 µm
  • Small spots behind a scanner and f-theta lens: LUM-F
  • Large beams after an expander, or large-area top-hats: LUM-B-L
  • Focus position together with depth of focus: LUM-Z

Full specifications are in each product page's datasheet.

5. Power and attenuation

Camera sensors saturate at very low power, so decide the attenuation first.

  • LUM-B-D: reflective ND filters (200–1064 nm, OD 0.1 up to 4.0)
  • LUM-B-N, LUM-B-M: absorptive ND filters (343–1064 nm, OD 0.1 up to 6.0)
  • LUM-B-L: input below 1 W recommended to protect the lens
  • LUM-Z: below 1 W

Higher powers need additional attenuation in front of the camera. Send us the laser power, wavelength and beam size and we will propose a configuration.

6. Measurement environment checklist

  • Measure after the laser has warmed up
  • Align the beam normal to the sensor
  • Dim room lighting and retake the background frame
  • Synchronize exposure with the trigger for pulsed lasers
  • Use a fixed mount where vibration is high
  • At high power, suspect thermal lensing in the attenuation optics themselves

7. Lumosity software and integration

Lumosity provides ISO 11146 analysis with 1D cross-section and 2D viewers and saves images and data. Its XML protocol over TCP/IP allows remote control and result transfer to the machine's PC software. Several LUM-B cameras can run in one Lumosity instance to monitor multiple laser heads on a production line.

8. Basic measurement procedure

Follow the steps in order.

  • Step 1 Check wavelength and power, choose the attenuation
  • Step 2 Center the beam on the sensor
  • Step 3 Turn the beam off and take a background frame
  • Step 4 Set exposure and ND so the peak sits just below saturation
  • Step 5 Record D4σ diameter, ellipticity and centroid; average repeated measurements if needed
  • Step 6 Compare with the reference to judge alignment and optics condition

If the model or setup is unclear, contact us with the laser specification and what you need to measure.