Understanding Laser Beam Quality and M²
- By
- Shinhotek LUMOS
- Published
- 2026-11-19
With the same power and the same lens, beam quality still changes the focused spot size. This guide explains what M² is and how it is measured.
What M² is
M² describes how much faster a real beam spreads than an ideal Gaussian (TEM00) beam.
- An ideal Gaussian beam has M² = 1
- Real beams are always above 1; the larger it is, the bigger the focused spot and the shorter the depth of focus with the same lens
- Single-mode lasers are usually close to 1, multi-mode lasers much higher
Relation to beam parameter product (BPP)
The product of waist radius w₀ and divergence half-angle θ is the beam parameter product.
- BPP = w₀ · θ = M² · λ / π
- At a given wavelength λ, M² and BPP are proportional
- Multi-mode beams such as high-power fiber lasers are often specified by BPP (mm·mrad) instead of M²
How M² affects the process
Focused spot size scales roughly with M², and Rayleigh length (depth of focus) scales inversely with it.
- Higher M² means a larger spot with the same lens and lower energy density
- Shorter depth of focus makes the process sensitive to part height
- M² often rises as a laser ages or optics get contaminated, so it is used as a maintenance indicator
Measuring per ISO 11146
M² cannot be obtained from a single cross-section; diameters are measured at several positions around focus.
- Focus the beam with a lens and measure the D4σ diameter at different axial positions
- ISO 11146 calls for at least 10 positions, half within one Rayleigh length of the waist and half beyond two Rayleigh lengths
- Fit the diameters to the hyperbolic propagation equation to obtain waist diameter, waist position, divergence and M²
- Fitting x and y separately also reveals astigmatism
Practical points
- Use D4σ at every position, with background subtraction and no saturation
- Choose magnification and sensor so even the waist spot covers 20 pixels or more
- Use a low-aberration focusing lens so its aberrations do not enter the result
- Positioning accuracy along the axis feeds directly into the result
Measuring around focus with LUMOS
- LUM-Z: travels up to 10 mm along the axis with a minimum 10 µm step and measures spot size, depth of focus and 3D intensity distribution; for beams larger than 3 µm, 343–1064 nm (selectable), below 1 W
- LUM-F: magnifies micrometre focused spots through an objective
- LUM-B: measures raw and expanded beam cross-sections
For a setup that needs M² values, send us the wavelength, power and beam size and we will advise.