Laser Beam Profiling for Semiconductor Processing
- By
- Shinhotek LUMOS
- Published
- 2026-11-19
In semiconductor laser processing, small changes in spot size and shape at the work plane move the yield. This note covers what to check per process, where to measure and how to monitor in production.
What to check per process
| Process | Typical beam | What to check |
|---|---|---|
| Wafer dicing, scribing | UV or ultrafast focused spot | Spot diameter, ellipticity, focus position |
| Marking | Scanner and f-theta focused spot | Spot size variation across the field |
| Drilling | Focused spot | Spot diameter and roundness |
| Annealing, lift-off | Line beam or top-hat | Plateau size, uniformity, edges |
When results drift without any process change, a change in spot shape is often the cause rather than laser power.
Where to measure
- Laser output (raw beam): the laser's own condition and mode, with the LUM-B series
- After the expander and optics: large expanded beams and top-hats; LUM-B-L for large sizes
- Work plane (focused spot): the spot that actually hits the part; micrometre spots are measured with LUM-F and its objectives, down to Ø 3 µm at ×20
- Around focus: when focus position and depth of focus set the process window; LUM-Z travels up to 10 mm along the axis and measures spot size, depth of focus and 3D intensity distribution
UV wavelengths
UV wavelengths common in dicing and micromachining make model choice important.
- LUM-B-D (LUM-B-DUV): 200–1064 nm, reflective ND filters for DUV/UV
- LUM-F-DUV: 240–355 nm
- LUM-F-UV: 343–355 nm
Focused spots for 355 nm dicing are measured with LUM-F-UV or LUM-F-DUV.
Production line monitoring
Measuring regularly under the same conditions catches problems earlier than a one-off check.
- Set reference values at a fixed position and exposure and compare diameter, ellipticity and centroid
- Connect several LUM-B cameras to one Lumosity instance to watch multiple laser heads
- Send results to the machine's PC software over Lumosity's XML protocol on TCP/IP for automatic pass/fail
Checklist
- Was the work-plane spot re-measured after optics replacement or alignment?
- Were spot sizes compared at the center and edge of the scanner field?
- Was the laser warmed up and the background frame subtracted?
- Was the sensor kept out of saturation?
- Are reference values and conditions (exposure, ND, position) recorded?
Contact
Tell us the process, wavelength, spot size and measurement position and we will propose a model and setup.