Gaussian and Top-hat Beam Measurement Comparison

By
Shinhotek LUMOS
Published
2026-11-19

Measured with the same beam profiler, Gaussian and top-hat beams still need different numbers. This guide compares their definitions, diameter conventions and measurement practice.

How the two beams differ

  • Gaussian beam: brightest at the center, intensity falls off smoothly outward. Most lasers emit this shape by default
  • Top-hat (flat-top) beam: flat intensity across the center with steep edges. Used where a large area needs the same energy, such as annealing, lift-off, welding and uniform illumination

Gaussian beams: which diameter

Gaussian beams have several diameter conventions, so state which one you use.

ConventionDefinitionRelation for a Gaussian
D4σ (ISO 11146)Second-moment diameterEquals the 1/e² diameter
1/e²Width at 13.5% of peakReference
FWHMWidth at 50% of peakAbout 0.59 × the 1/e² diameter

For a Gaussian a single D4σ value represents size, which makes comparison easy. Adding ellipticity and centroid drift tells you about alignment.

Top-hat beams: one diameter is not enough

For a top-hat, D4σ differs from the plateau size and may barely change even when the shape degrades. Check three things together:

  • Plateau size: width of the region above a set fraction of peak (e.g. 90%), from a threshold diameter
  • Uniformity (flatness): how even the intensity is across the plateau, seen in the cross-section
  • Edge steepness: how quickly intensity drops at the plateau edge; narrower edges give sharper process boundaries

Measurement tips

ItemGaussianTop-hat
ExposurePeak just below saturationWhole plateau below saturation
ResolutionDiameter of 20 pixels or moreEdge spans several pixels
BackgroundMust be subtracted; tails are sensitiveRemove noise outside the plateau
View2D view and D4σ1D cross-section and threshold diameter

With a wide plateau, partial saturation is easy to miss on screen. Always check whether the top of the cross-section has been clipped flat.

Large top-hats: sensor size comes first

Top-hats range from a few millimetres to tens of millimetres, and the whole beam must fit on the sensor to judge uniformity.

  • LUMOS IFI optics turn fiber-delivered lasers into square flat-top beams from 2 × 2 mm up to 30 × 30 mm, depending on the model
  • Such large top-hats can be measured in one shot with LUM-B-L and its 38.75 × 27.75 mm sensor
  • Top-hats of a few millimetres or less are measured with LUM-B

Summary

For Gaussian beams look at D4σ diameter and ellipticity; for top-hats look at plateau size, uniformity and edge steepness. Tell us the beam size and wavelength and we will recommend a model and setup.